Technical data
SERVice Subsystem 16
N1911A/1912A P-Series Power Meters Programming Guide 707
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the rear panel trig out 
for testing purposes.
If the command is set to:
• ON, the 10 MHz time base signal is sent to the rear panel trigger out 
connector.
• OFF, the 10 MHz time base signal is disabled. 
Syntax
Example
NOTE
This command overrides the OUTPut:TRIGger[:STATe] command. 
For example, if OUTPut:TRIGger[:STATe] is ON and the command 
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets it to 
OFF. However, the 10 MHz remains out the Trig out port. 
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the 
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now routed to 
the Trig out overriding the service command turning the 10 MHz to off.
SERV:BIST:TBAS:STAT OFF This command disables the signal. 
:BIST
SERV
0|OFF
1|ON
Space
?
:TBAS
:STAT