Technical data
16 SERVice Subsystem
724 N1911A/1912A P-Series Power Meters Programming Guide
SERVice:SENSor[1]|2:PCALfactor <cal_factor_data> 
This command writes calibration factor data to, or reads calibration factor 
data from, the currently connected sensor. The whole calibration factor 
block must be written at once as a checksum is generated. The new block 
must not be larger than the existing block.
This command applies to E9320 Series sensors for peak path data only. 
For E4410 Series, E9300 Series and E9320 Series sensors, average path 
data, refer to “SERVice:SENSor[1]|2:CALFactor <cal_factor_data>” on 
page 716.
Syntax
Parameters
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
Item Description/Default
cal_factor_data A binary data block. Refer to Appendix A, “Calibration Factor Block 
Layout,” on page A-763 for further information. 