16 SERVice Subsystem
726 N1911A/1912A P-Series Power Meters Programming Guide
SERVice:SENSor[1]|2:POWer:AVERage:MAXimum? 
This query returns the maximum average power that can be measured by 
the currently connected sensor. It is applicable to E- Series sensors only. 
Maximum average power information is stored in the sensor’s EEPROM.
Syntax
Example
Error Messages
• If no sensor is connected, error –241, “Hardware missing” occurs.
• If a a sensor other than an E- Series sensor is connected, error –241 
“Hardware missing” occurs.
• If the E- Series sensor currently connected does not contain the 
necessary information in EEPROM, error –241 “Hardware missing” 
occurs.
SERV:SENS:POW:AVER:MAX? This query returns the maximum average 
power that can be measured by the 
E- Series sensor currently connected to 
Channel A.
:POW
?
:AVER
:MAX
SERV
1
2
:SENS