Technical data
Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 51
Figure 1-7 Generalized Status Register Model
When a status group is implemented in an instrument, it always contains 
all of the component registers. However, there is not always a 
corresponding command to read or write to every register.
Condition Register
The condition register continuously monitors the hardware and firmware 
status of the power meter. There is no latching or buffering for this 
register, it is updated in real time. Condition registers are read- only.
Transition Filter
The transition filter specifies which types of bit state changes in the 
condition registers and set corresponding bits in the event register. 
Transition filter bits may be set for positive transitions (PTR), negative 
transitions (NTR), or both. Transition filters are read- write. They are 
unaffected by *CLS or queries. After STATus:PRESet the NTR register is 
set to 0 and all bits of the PTR are set to 1.
Event Register
The event register latches transition events from the condition register as 
specified by the transition filter. Bits in the event register are latched and 
on setting they remain set until cleared by a query or a *CLS. Also on 
setting, an event bit is no longer affected by condition changes. It remains 
set until the event register is cleared; either when you read the register or 
when you send the *CLS (clear status) command. Event registers are 
read- only.
0
1
2
Bit 0
Bit 1
Bit 2
Bit 3
Condition
Register
Tr an sit io n
Filter
Event
Register
Enable
Register
Logical OR
Summary
Bit