Technical data
Calibration Factor Block Layout A
N1911A/1912A P-Series Power Meters Programming Guide A-767
Number of tables 1 1 - - None Number of cal factor 
tables. This is currently 
unused but has been 
set to a default value of 
1.
Number of frequency 
points
2 - 16 bit 
integer
-None
Bytes per 
frequency 
point
1 - - - None Number of bytes in 
cal factor value at 
each 
frequency.
Frequency LSB weight 2 1000 - Hertz Fhbp (Freq. Hz per 
bit). 
1 KHz per bit for the 
cal factor: 1 KHz x 
2^32 = 4.3E+12 = 
4300 GHz range
Header Total: 6
Cal Factor Table:
Frequency (point ā0ā) 4 - 32 bit fixed 0 to Fhpb* 
(2^32)
None Fhpb = Freq Hz per bit
Cal factor
2
2 - 2.14 0.25 to 3 None Used to adjust analog to 
digital convertor (ADC) 
values.
These table entries are repeated as shown for each frequency point
Frequency (point āNā) 4 - 32 bit fixed 0 to Fhpb* 
(2^32)
None Fhbp = Freq Hz per bit
Cal factor 2 - 2.14 0.25 to 3 None Used to adjust analog to 
digital converter (ADC) 
values.
Tab le  Si ze:
-
See note
3
The table size is 
dependent on the 
number of frequency 
points.
E9320 Series 
Sensors: Calibration 
Factor Block Layout
No. 
Bytes
Contents Data 
Format
Data 
Range
Units Notes