Device Status Register
The device status register set contains bits which give device dependent 
information.
The following bits in these registers are used by the power meter:
Tab le 1- 11 Bit Definitions - Device Status Register
11 Lower Limit 
Fail
This is a summary bit for the Lower Limit Fail Register.
• SET: If a measurement is made and either a channel or window lower limit test 
fails.
• CLEARED: If a measurement is made and the lower limit test is not enabled or 
the test is enabled and passes.
12 Upper Limit 
Fail
This is a summary bit for the Upper Limit Fail Register.
• SET: If a measurement is made and either a channel or window upper limit test 
fails.
• CLEARED: If a measurement is made and the upper limit test is not enabled or 
the test is enabled and passes.
Bit 
Number
Meaning EVENts Causing Bit Changes
Bit 
Number
Decimal 
Weight
Definition
0 - Not used
1 2 Channel A sensor connected 
24
Channel B sensor connected
*
3 8 Channel A sensor error
416
Channel B sensor error
*
5 32 Channel A sensor Front/Rear
664
Channel B sensor Front/Rear
*
14 16384 Front Panel key press
* N1912A only