Technical data
1 Power Meter Remote Operation
66 N1911A/1912A P-Series Power Meters Programming Guide
The condition bits are set and cleared under the following conditions:
Tab le 1- 12 Bit change conditions for Device Status Register
Bit 
Number
Meaning EVENts Causing Bit Changes
1Channel A 
sensor 
connected
• SET: When a power sensor is connected to the Channel A 
input.
• CLEARED: When no power sensor is connected to the 
Channel A input.
2Channel B 
sensor 
connected
• SET: When a power sensor is connected to the Channel B 
input.
• CLEARED: When no power sensor is connected to the 
Channel B input.
3Channel A 
error
• SET: If the power sensor EEPROM on Channel A has failed 
or if there are power sensors connected to both the rear 
and front panel Channel A connectors.
• CLEARED: In every other condition.
4Channel B 
error
• SET: If the power sensor EEPROM on Channel B has failed 
or if there are power sensors connected to both the rear 
and front panel Channel B connectors.
• CLEARED: In every other condition.
5Channel A 
Front/Rear
• SET: If a power sensor is connected to the Channel A rear 
panel.
• CLEARED: If a power sensor is connected to the Channel A 
front panel.
6Channel B 
Front/Rear
• SET: If a power sensor is connected to the Channel B rear 
panel.
• CLEARED: If a power sensor is connected to the Channel B 
front panel.
14 Front Panel 
Key Press
This is an event, and DOES NOT set the condition register. The 
bit is set in the event register which is cleared when read. Note 
that the transition registers are of no use for this bit.