Technical data
19
The wide-dynamic square-law range 
of the U2000 Series sensors makes 
them ideally suited to measure signal 
formats with complex modulation 
formats which cause high crest 
factors. The principle is that the 
peaks of the crest factor spikes do 
not exceed the square-law detection 
range. Crest factors of ten are 
easily handled by assuring that the 
measurement range of the sensor is 
dialed down in such a way that the 
peak power does not exceed +20 dBm. 
For users measuring complex power 
formats with high crest factors, it is 
recommended that they refresh their 
background on signal formats on the 
use of dual path sensors by referring 
to [Ref: 2, pages 29 – 30] [Ref: 2, pages 
3 – 6].
Applications that require power 
measurements of pulsed RF/
microwave signals (such as radar 
signals) are covered in the U2000 
Series too, provided that it is permitted 
to just measure the average power 
of the pulsed modulation waveform. 
In many if not most of the system 
tests, pulsed power transmitters have 
fi xed duty cycle (peak power duration 
divided by pulsed repetition period), 
thus, an average power reading can be 
computed to peak power. Again, for 
proper averaging, it must be assured 
that the peak pulse power entering 
the U2000 Series does not exceed +20 
dBm so as not to get out of square-law 
range. This may require an external 
microwave attenuator pad of 10 dB or 
20 dB.
The calibration factor versus frequency 
versus temperature characterization is 
done once at the factory during 
manufacturing to measure each 
individual diode stack assembly for 
temperature variations. The data that 
is run against a frequency response 
test forms the basis for the
3-dimensional correction table stored 
on-board within the U2000 Series in 
the SDRAM and Flashdrive. [Ref: 5] 
The test of the stability of the 
correction table over several 
time periods has resulted in 
recommendations that re-calibration 
is only required once per year. The 
sensor works within a specifi ed 
warranty period of one year. 
Furthermore, when the yearly 
calibration cycle is due, the calibration 
facility only needs to make a 
frequency response test at laboratory 
temperature (25°C). It has been 
determined that the temperature 
sensitivity profi le stays constant over 
time as long as no overload damage 
has been experienced. The crucial 
point of remote installations of the 
U2000 Series is the capability to 
perform automatic zeroing without 
having to disconnect the RF connector 
at the power sensing point, or shut 
down the system power. This function 
is described as an integrated function 
on-board the same microcircuit that 
holds the diode stacks, and the signal 
switching from high to low sensitivity 
channels. [Ref: 5]
The calibration process may be 
performed by returning the U2000 
Series to the Agilent Service Facility 
or it can be done in the user’s Cal 
Lab. Once the new correction data is 
obtained, instructions are available to 
permit the calibration facility to 
input new data tables into the 
individual sensor via the USB 
communications bus. 
Signal Formats
Calibration, Zero and Cal










