Technical data
Repair Guide 6
N1911A/1912A P-Series Power Meters Service Guide 99
Step 2:
• Use ODU socket to remove N1912-61806 
sensor flex assembly
• Release the tab holding the flex cable to the 
display interface board, and then disconnect 
it
Step 3:
• Disconnect the white plastic plug from the 
display interface board
Step 4:
• Unlock the main plastic clip that holds the 
front panel sub-frame and display support 
molding together (situated beside the key 
flex circuit), and carefully pull them apart to 
separate them
Instructions Visual
Flex Cable
White Plastic Plug