Technical data
Troubleshooting Guide 5
N1911A/1912A P-Series Power Meters Service Guide 53
ChB peak path Verifies that the peak path 
of channel B is working
(Note: This does not prove 
that the peak path meets its 
specifications)
Replace the DAP assembly 
for channel B to see if this 
clears the fault
DAP assembly, channel B 
(low probability)
Main board (high probability)
ChB CW path Verifies that the average 
path of channel B is 
working
(Note: This does not prove 
that the average path 
meets its specifications)
Not applicable Main board
ChB DAP check Executes an internal self-test 
procedure on the DAP 
assembly for channel B
Replace the DAP assembly 
for channel B to see if this 
clears the fault
DAP assembly, channel B 
(high probability)
Main board (low probability)