Technical data
Repair Guide 6
N1911A/1912A P-Series Power Meters Service Guide 73
Rear Panel Assembly
Main Assembly
Customization Details
The standard P- Series power meters have the input sensor connector(s) and 
reference calibrator connector on the front panel. Option 003 is available 
to move the input sensor connector(s) and reference calibrator connector 
to the rear panel. Below are the customization details on rear panel assem-
bly.
Keysight Part 
Number
Description Visual
N1912-61007
N1912-61031
Rear panel assembly [new]
Revised rear panel assembly [new]
Note:
• The rear panel assembly must be 
customized to suit the hardware 
configuration of the unit being repaired
• Refurbished rear panel assemblies are not 
available
• Parts N1912-61007 and N1912- 61031 differ 
slightly with regards to the connector 
cutout positions
• The revision of the main board can be used 
to help determine which rear panel 
assembly is in use (see “Additional Repair 
Notes” on page 108)
• The same assemblies are used for both 
N1911A and N1912A models
N1912-61007 / N1912-61031
Connector Option Details Part Number
N1911A (Front connectors option)
1 unit of rear panel plug (BNC)
2 unit of rear panel plug (Sensor)
1 unit of rear panel plug (Calibrator)
6960-0081
6960-0024
6960-0178