General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 81
4 Press . The Trigger menu is displayed.
5 Press   softkey to configure a trigger. 
6 Select either   or  .
7 Press   to configure the remaining trigger parameters. The trigger 
Settings menu consists of two pages. Figure 2- 61 shows page 1 and 
Figure 2- 62 shows page 2.
8 Press   and   will be automatically enabled.
9 Press  , select   to allow a TTL level high to be produced 
at the rear panel TRIG OUT BNC port upon the meter completed a 
measurement. Refer to note below for details.
NOTE
When E-Series E9320 or P-Series N1920 sensor is connected, the default sensor mode is 
Normal. User can use this Normal mode to perform power or frequency sweep, or change 
the mode to AVG only to perform power or frequency sweep. Refer Ta b l e 2 - 6  on the 
differences in measurement functions supported under these two modes.
To change the sensor mode from the default Normal mode to AVG only mode, follow the 
instructions below:
• Press  . Use the   and   keys to highlight the Sensor Mode field.
• Press   to display the Sensor Mode pop-up. 
• Use the   and   keys to highlight the AVG and press  .
Acqn
Sing Trig
Cont Trig
Settings
Source
Ext
1 of 2
Output On
NOTE
In both power sweep and frequency sweep mode, the meter TRIG OUT to power source 
TRIG IN connection is optional. If this connection was not setup previously, you need to 
setup the proper Dwell time. You may choose to setup the Dwell time in the power source 
step setting to cater the maximum settling time required by the power meter. However, to 
achieve maximum speed in frequency sweep, this setup is recommended. Once the power 
reading is settled, the meter will output a TTL signal at the TRIG OUT to the power source 
TRIG IN to trigger the power source to step to the next point.