User`s guide
General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 85
Determine the Right Step to be Set 
Number of frequency step can be calculated by using equation below:
Step = f
stop
 – f
start
 + Interval
  Interval
where,
Step = Number of frequency step
f
start
 = Frequency sweep’s start point
f
stop
 = Frequency sweep’s stop point 
Interval = Frequency step size
Example
When f
start
 = 1 GHz and f
stop
 = 5 GHz with given interval of 0.5 GHz, the 
Step
should be set to
Step
= f
stop
 – f
start
 + Interval
  Interval
= 5 GHz – 1 GHz + 0.5 GHz 
   0.5 GHz
= 9