Recorder Output
The rear panel Recorder Output connectors (1 and 2) produce a dc 
voltage that corresponds to the power level in Watts of the channel 
depending on the measurement mode. This dc voltage ranges from 0 to 
+1 Vdc. The output impedance is typically 1 kΩ. Channel and display 
offsets, and duty cycle have no effect on the Recorder Outputs.
For example, the Recorder Outputs can be used to:
• Record swept measurements
• Level an output from a source using external leveling 
• Monitor the output power
To access the Recorder menu press  , and enable Rec o/p. This 
allows you to switch the Recorder Output signal either on or off. The 
 and   softkeys allow you to scale the power levels to 
represent the 1 Vdc maximum and 0 Vdc minimum output voltage of the 
Recorder Output.
Procedure
Set the recorder output as follows:
1 Press  ,   to display the Measurement Setup menus. 
2 Use the   and   keys to highlight the Rec o/p: setting field.
3 Press   to reveal the Recorder option menu.
Max Power
Min Power
NOTE
Ensure you have selected the channel you want to set up.
Meas Select