User`s guide
Using E9320 E-Series Power Sensors 4
N1911A/1912A P-Series Power Meters User’s Guide 139
Step 1. Connecting the Power Sensor
When you initially connect an E- Series E9320 power sensor to the power 
meter, the sensor’s calibration data is automatically read by the power 
meter. After every power cycle, the E9320 must perform zero and 
calibration as the P- Series power meters do not retain the states and 
settings upon power cycle (warm start feature).
Connect the sensor to the power source.
Step 2. Setting the Channel Frequency
• Press . The Channel Setup screen is displayed. If using an 
N1912A, select the channel you want to configure. 
• Use the  ,  ,  , and   keys to highlight the frequency field. Press
 to configure the setting. See “Setting the Measurement 
Frequency” on page 32 for further information.
When you have completed the procedure you can return to this menu later 
to add more accuracy to the measurement.
Step 3. Setting the Trigger
To use the measurement gates the power meter must be triggered.
A trigger can be taken from a rising or falling measured power level or 
controlled externally using the Ext Trig input.
• Press . The Trigger menu is displayed. 
The trigger status is displayed below the   label in the Trigger 
menu. Press the   softkey and select   to configure 
continuous triggering. 
When you have completed the procedure you may return to the Trigger 
menu later to add more accuracy to the remaining trigger parameters, 
to help you achieve a stable and reliable trigger. 
Acqn 
Acqn 
Cont Trig