5 Using E9300 E-Series Power Sensors
156 N1911A/1912A P-Series Power Meters User’s Guide
4 Press   and the channel   softkey to start the calibration 
routine. The Calibrating pop- up is then displayed.
Now set the frequency of the signal you want to measure. The power 
meter automatically selects the appropriate calibration factor.
5 Press  . On dual channel meters select the required channel.
6 Use the   and   keys to highlight the Frequency value field and press
 to display the Frequency pop- up. Use the numeric keypad to 
enter the required value in the Frequency pop- up window.
Figure 5-115Frequency pop-up
7 Confirm your choice by pressing   or  .
8 Press   key to close the Channel Setup screen.
Now make the measurement.
9 Reconnect any required attenuators or adaptors and connect the power 
sensor to the signal to be measured.
The corrected measurement result is displayed.
Cal
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as 
follows:
1 Connect the power sensor to the POWER REF output.
2 Press   and  . (For dual channel meters, press 
,  or   as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
MHz
GHz