Achieving Stable Results with GSM Signals
Signals with a pulse repetition frequency (PRF) close to a multiple or 
sub- multiple of the 440 Hz chopper- amplifier signal generate a beat note 
at a frequency between the PRF and 440 Hz. Control over the filter 
settings is again required to obtain stable results. 
Enter
NOTE
You should also ensure the filter is not reset when a step increase or decrease in power is 
detected by switching the step detection off.
TIP The PRF of a GSM signal is approximately 217 Hz and thus 
requires more averaging than most other TDMA signals. To 
achieve a stable measurement use the filter setting procedures 
to set the  . Experimentally, a   setting of 148 
gives optimum results although settings in the order of 31 or 32 
give acceptable results if a faster measurement is required.