Measurement Accuracy and Speed 
The power meter has no internal ranges. The only ranges you can set are 
those of the E9300 E- Series power sensors (and other Agilent 
Technologies E- Series power sensors). With an E- Series E9300 power 
sensor  the range can be set either automatically or manually. Use 
autoranging when you are not sure of the power level you are about to 
measure. 
Setting the Range
There are two manual settings, LOWER  and  UPPER. The LOWER range 
uses the more sensitive path and the UPPER range uses the attenuated 
path in the E9300 E- Series power sensors.
The default is AUTO. In AUTO the range crossover value depends on the 
sensor model being used.
CAUTION
To prevent damage to your sensor do not exceed the power levels specified in the sensor 
user’s guide.
The 
 E9304A sensor is DC coupled. DC voltages in excess of the maximum value (5 Vdc) 
can damage the sensing diode.
Sensor LOWER range UPPER range
E9300/1/4A –60 dBm to –10 dBm –10 dBm to +20 dBm
E9300/1B –30 dBm to +20 dBm +20 dBm to +44 dBm
E9300/1H –50 dBm to 0 dBm 0 dBm to +30 dBm
E9300/1/4A E9300/1B E9300/1H
–10 dBm ±0.5 dBm +20 dBm ±0.5 dBm 0 dBm ±0.5 dBm