5 Using E9300 E-Series Power Sensors
164 N1911A/1912A P-Series Power Meters User’s Guide
Procedure
Set the range as follows:
1 Press 
. On dual channel meters, select the required channel. 
2 Use the   and   keys to select the Range: setting field.
3 Press   to display the Range pop- up.
4 Use the   and   keys to select the required setting.
5 Press   to complete the entry.
Measurement Considerations
While autoranging is a good starting point, it is not ideal for all 
measurements. Signal conditions such as crest factor or duty cycle may 
cause the power meter to select a range which is not the optimum 
configuration for your specific measurement needs. Signals with average 
power levels close to the range switch point require you to consider your 
needs for measurement accuracy and speed. For example, using an 
E9300/1/4A sensor, where the range switch point is –10 ± 0.5 dBm in a 
pulsed signal configured as follows: 
The calculated average power is –12 dBm.
Characteristic Value
Peak Amplitude –6 dBm
Duty Cycle 25%