Measurement Accuracy
Power sensors have small errors in their response over frequency. The 
response of each sensor is measured during manufacture (and during 
periodic calibration). With E- Series power sensors, the resulting frequency 
compensation information is written into Electrically Eraseable 
Programmable Read Only Memory (EEPROM). This allows the frequency 
and calibration data to be downloaded to the power meter automatically.
Using calibration factors enables you to achieve improved measurement 
accuracy. This section describes making continuous wave measurements 
using the E4410 E- Series power sensors.
Making a measurement requires the following steps:
1 Zero and calibrate the power meter/sensor combination.
2 Set the frequency for the signal you want to measure.
3 Make the measurement.
Procedure
First zero and calibrate the power meter/sensor combination.
1 Ensure the power sensor is disconnected from any signal source.
2 Press   and the channel   softkey. The Zeroing pop- up is 
displayed. 
3 Connect the power sensor to the POWER REF output.
4 Press   and the channel   softkey to start the calibration 
routine. The Calibrating pop- up is then displayed.
Zero
Cal