Figure 7-125Reference Calibration Factor pop-up window
5 Press 
 to complete the entry.
Now zero and calibrate the power meter/sensor combination as follows:
6 Press   and the channel   softkey. The Zeroing pop- up is 
displayed. 
7 Connect the power sensor to the POWER REF output.
8 Press   and the channel   softkey to start the calibration 
routine. The Calibrating pop- up is then displayed.
Now set the sensor calibration factor for the frequency of the signal you 
want to measure.
9 Check the current reference calibration factor setting by pressing 
,  . The value is displayed on the Cal Fac field.
%
Zero
Cal
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as 
follows:
1 Connect the power sensor to the POWER REF output.
2 Press   and  . (For dual channel meters, press 
,  or   as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
Offset