User`s guide
Using 8480 Series Power Sensors 7
N1911A/1912A P-Series Power Meters User’s Guide 183
Example
To make a measurement on channel A with a power sensor which has a 
reference calibration factor of 99.8% and a calibration factor of 97.6% at 
the measurement frequency.
• Disconnect the power sensor from any signal source.
• Press , 
 and the channel   softkey.
• Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop- up 
window.
• Press 
 to complete the entry.
• Press   and the channel   softkey. The Zeroing pop- up is 
displayed. 
• Connect the power sensor to the POWER REF output.
• Press   and the channel   softkey to start the calibration 
routine. The Calibrating pop- up is then displayed.
• Press  ,  . The value is displayed on the Cal Fac field.
• Use the   and   keys to highlight the Cal Fac value field and press 
 to display the Cal Factor pop- up. Use the numeric keypad to enter 
97.6 in the Cal Factor pop- up window.
• Press 
 to complete the entry.
• Connect the power sensor to the signal to be measured.
• The corrected measurement result is displayed.
REF CFs
REF CF 
%
Zero
Cal
Offset
%
NOTE
When no sensor tables are selected and  display mode is chosen, the 
calibration factor used for the measurement is displayed in the upper window as shown in 
Figure 7-127.
Single Numeric