User`s guide
Using N8480 Series Power Sensors 8
N1911A/1912A P-Series Power Meters User’s Guide 205
Now set the sensor calibration factor for the frequency of the signal you 
want to measure.
9 Check the current reference calibration factor setting by pressing 
,  . The value is displayed on the Cal Fac field.
Does this setting match the value for the sensor? (The power sensor 
reference calibration factor can normally be found above the calibration 
factors table on the power sensor body.)
10 To change the settings, use the   and   keys to highlight the Cal Fac
value field and press   to display the Cal Factor pop- up. Use the 
numeric keypad to enter the required value in the Cal Factor pop- up 
window.
Figure 8-139Calibration factor pop-up window
11 Press 
 to complete the entry.
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as 
follows:
1 Connect the power sensor to the POWER REF output.
2 Press   and  . (For dual channel meters, press 
,  or   as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
Offset
%