User`s guide
Using N8480 Series Power Sensors 8
N1911A/1912A P-Series Power Meters User’s Guide 215
4 Press   to complete the entry.
Enter (or edit) the frequency and  cal factor pairs as follows:
1 Press   to add a new frequency value (or press   to edit). 
Use the numeric keypad to enter the required value in the Frequency 
pop- up window. Complete the entry by pressing the  ,   keys.
2 Enter the new cal factor value (or press   to edit). Use the 
numeric keypad to enter the required value in the Cal Factor pop- up 
window. Complete the entry by pressing the   key. 
3 Continue adding/editing values until you have entered all the data you 
require.
4 When you have finished editing the table press   to save the table.
Enter
NOTE
A calibration factor in the range of 1 % to 150 % can be entered. 
The following rules apply to naming sensor calibration tables:
• The name must consist of no more than 12 characters.
• All characters must be upper or lower case alphabetic characters, or numeric (0-9), or 
an underscore (_).
• No other characters are allowed.
• No spaces are allowed in the name.
Insert
Change
GHz
MHz
Change
%
Done
NOTE
Ensure that the frequency points you use cover the frequency range of the signals you want 
to measure. If you measure a signal with a frequency outside the frequency range defined 
in the sensor calibration table, the power meter uses the highest or lowest frequency point 
in the sensor calibration table to calculate the offset










