User`s guide
9CCDF
224 N1911A/1912A P-Series Power Meters User’s Guide
While the marker is moving along the trace, the power (dB) and the 
probability (%) will be reflected at the lower left corner as shown in 
Figure 9- 151. Row 1 shows the reading for Marker 1 while row 2 shows 
the reading for Marker 2. The lower right corner shows the delta reading 
for Marker 2- Marker 1, average measurement, peak measurement and the 
peak- to- average ratio for the respective channel that you have selected. 
Step 3: Setting the Scale/Div
• You are allowed to optimize the view of the trace by setting the 
appropriate scale/div. The lowest limit is 0.1 dB while the highest limit 
is 50 dB. A warning message will appear if the lower or upper limit has 
been exceeded.
NOTE
The alphabet in the parenthesis after Average, Peak and Peak-Average measurements 
represents: (A) - Channel A, (B) - Channel B, (G) - Gaussian, and (R) - Reference.
NOTE
Press   once to view the displayed trace in full-screen.
NOTE
Marker control menu is also accessible in the Trace  C o n trol menu.