User`s guide
2 General Power Meter Functions
48 N1911A/1912A P-Series Power Meters User’s Guide
Enter (or edit) the frequency and offset pairs as follows:
1 Press   to add a new frequency value (or press   to edit). 
Use the numeric keypad to enter the required value in the Frequency 
pop- up window. Complete the entry by pressing the  ,   keys.
2 Enter the new offset value (or press   to edit). Use the numeric 
keypad to enter the required value in the Offset pop- up window. 
Complete the entry by pressing the   key. 
3 Continue adding/editing values until you have entered all the data you 
require.
4 When you have finished editing the table press   to save the table.
Insert
Change
GHz
MHz
Change
%
Done
NOTE
If you measure a signal with a frequency outside the frequency range defined in the 
frequency dependent offset table, the power meter uses the highest or lowest frequency 
point in the frequency dependent offset table to calculate the offset