User`s guide
2 General Power Meter Functions
68 N1911A/1912A P-Series Power Meters User’s Guide
Mode The   key is only available when trigger   is selected. 
To change the setting press   and choose   or  . 
The current setting is displayed below the label. When   is selected 
you can choose the RF power level transition used as the trigger. The 
power meter automatically finds a triggering power level transition when 
 is selected.
Level  is only available when   trigger mode has been selected. 
The current value is shown below the   label. The minimum power 
level you can enter is limited to 40 dB below the maximum sensor power. 
To change the setting press   and use the numeric keypad to enter 
the required value in the Trigger Level pop- up window to enter the new 
value.
Figure 2-53 Trigger Level pop-up
Complete the entry by pressing  .
Delay  The current setting is displayed below the   label. The delay 
time is applied between the trigger event and all the gate start times. This 
allows you to time- shift all the gates by the same amount with one setting 
change. A delay of up to 1 second can be entered. To enter or change the 
setting press   and use the numeric keypad to enter the required 
value in the Trigger Delay pop- up window to enter the new value.
NOTE
For N1912A, by pressing  , you can choose your source through the three 
different softkeys:  ,  , and  .
Source
Ext
Ch.A
Ch.B
Mode
Source Int
Mode
Norm
AutoLvl
Norm
AutoLvl
Level
Norm
Level
Level
dBm
Delay
Delay