User`s guide
General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 71
Falling Edge
When a falling power transition triggers the power meter, the triggering 
system is disabled. The power meter does not trigger again if another 
falling power transition is presented. The triggering system is rearmed 
only when the input power rises above a level equal to the trigger level 
plus the configured hysteresis value.
To change the setting, press   and use the numeric keypad to 
enter the required value in the Trigger Hysteresis pop- up window to enter 
the new value.
Figure 2-57 Trigger Hysteresis pop-up
Complete the entry by pressing  .
Output The current setting is displayed below the   label. A TTL 
level high is produced at the rear panel TRIG OUT BNC connector when the 
power meter is triggered when   is configured.
To change the setting, press   to highlight   or   as required.
Hysteresis
dB
Output
On
Output
On
Off