2 General Power Meter Functions
72 N1911A/1912A P-Series Power Meters User’s Guide
Setting External Trigger for Normal and Average Power 
Measurement
There are two modes featured in peak and average power measurement
• Power sweep mode
• Frequency sweep mode
These modes are used to eliminate the need for lengthy test routines, 
while increasing measurement throughput by reducing overhead of 
communication with the controller. The sweep feature allows you to make 
power measurement by quickly stepping through a series of frequencies or 
power levels. Configuration shown in Figure 2- 58 illustrates the triggering 
connection required to synchronize power meter measurement to the 
power source settings.
Table 2- 5 below shows the power and frequency sweep mode compatibility 
in normal and average mode. Measurement averaging and trace display 
will be turned OFF in E- Series E9320 and P- Series N1920 power sensors 
while performing power or frequency sweep during normal mode.
Table 2 - 5 Power and frequency sweep mode compatibility table
* Average only mode measurement is calculated based on peak measurement operation 
for the P-Series N1920 power sensor.
Sensor Normal Mode Average Mode
848x  —
✔
E441x  —
✔
E930x  —
✔
E932x
✔✔
N848x  —
✔
N192x
✔
✔
(emulated*)
NOTE
Auto CAL will be temporary disabled in P-Series N1920  power sensors during power or 
frequency sweep.