Technical data
s1
147
Agilent InfiniiVision 2000 X-Series Oscilloscopes 
Programmer's Guide
6
Root (:) Commands
Control many of the basic functions of the oscilloscope and reside at the 
root level of the command tree. See "Introduction to Root (:) Commands" 
on page 150.
Table 49 Root (:) Commands Summary
Command Query Options and Query Returns
:ACTivity (see 
page 151)
:ACTivity? (see 
page 151)
<return value> ::= 
<edges>,<levels>
<edges> ::= presence of edges 
(32-bit integer in NR1 format)
<levels> ::= logical highs or 
lows (32-bit integer in NR1 
format)
n/a :AER? (see page 152) {0 | 1}; an integer in NR1 format
:AUToscale 
[<source>[,..,<source
>]] (see page 153)
n/a <source> ::= CHANnel<n> for DSO 
models
<source> ::= {CHANnel<n> | 
DIGital<d> | POD1 | POD2} for MSO 
models
<source> can be repeated up to 5 
times
<n> ::= 1 to (# analog channels) 
in NR1 format
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
:AUToscale:AMODE 
<value> (see 
page 155)
:AUToscale:AMODE? 
(see page 155)
<value> ::= {NORMal | CURRent}}
:AUToscale:CHANnels 
<value> (see 
page 156)
:AUToscale:CHANnels? 
(see page 156)
<value> ::= {ALL | DISPlayed}}
:AUToscale:FDEBug {{0 
| OFF} | {1 | ON}} 
(see page 157)
:AUToscale:FDEBug? 
(see page 157)
{0 | 1}