:DIGital<d> Commands
Control all oscilloscope functions associated with individual digital 
channels. See "Introduction to :DIGital<d> Commands" on page 243.
Introduction to
:DIGital<d>
Commands
<d> ::= 0 to (# digital channels - 1) in NR1 format
Table 62 :DIGital<d> Commands Summary
Command Query Options and Query Returns
:DIGital<d>:DISPlay 
{{0 | OFF} | {1 | 
ON}} (see page 245)
:DIGital<d>:DISPlay? 
(see page 245)
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
{0 | 1}
:DIGital<d>:LABel 
<string> (see 
page 246)
:DIGital<d>:LABel? 
(see page 246)
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<string> ::= any series of 10 or 
less ASCII characters enclosed in 
quotation marks
:DIGital<d>:POSition 
<position> (see 
page 247)
:DIGital<d>:POSition? 
(see page 247)
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<position> ::= 0-7 if display 
size = large, 0-15 if size = 
medium, 0-31 if size = small
Returns -1 when there is no space 
to display the digital waveform.
:DIGital<d>:SIZE 
<value> (see 
page 248)
:DIGital<d>:SIZE? 
(see page 248)
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<value> ::= {SMALl | MEDium | 
LARGe}
:DIGital<d>:THReshold 
<value>[suffix] (see 
page 249)
:DIGital<d>:THReshold
? (see page 249)
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<value> ::= {CMOS | ECL | TTL | 
<user defined value>}
<user defined value> ::= value in 
NR3 format from -8.00 to +8.00