19 :MARKer Commands
:MARKer:XUNits:USE
 (see page 790)
Command Syntax
:MARKer:XUNits:USE
When DEGRees is selected for :MARKer:XUNits, the :MARKer:XUNits:USE 
command sets the current X1 location as 0 degrees and the current X2 
location as 360 degrees.
When PERCent is selected for :MARKer:XUNits, the :MARKer:XUNits:USE 
command sets the current X1 location as 0 percent and the current X2 
location as 100 percent.
Once the 0 and 360 degree or 0 and 100 percent locations are set, inputs 
to and outputs from the :MARKer:X1Position, :MARKer:X2Position, and 
:MARKer:XDELta commands/queries are relative to the set locations.
See Also • "Introduction to :MARKer Commands" on page 318
• ":MARKer:XUNits"  on page 325
• ":MARKer:X1Y1source"  on page 321
• ":MARKer:X2Y2source"  on page 323
• ":MEASure:SOURce"  on page 361
• ":MARKer:X1Position"  on page 320
• ":MARKer:X2Position"  on page 322
• ":MARKer:XDELta"  on page 324