Technical data
:MEASure Commands 20
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 335
:MEASure:NWIDth 
[<source>] (see 
page 352)
:MEASure:NWIDth? 
[<source>] (see 
page 352)
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>} for 
DSO models
<source> ::= {CHANnel<n> | 
DIGital<d> | FUNCtion | MATH | 
WMEMory<r>} for MSO models
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<return_value> ::= negative 
pulse width in seconds-NR3 format
:MEASure:OVERshoot 
[<source>] (see 
page 353)
:MEASure:OVERshoot? 
[<source>] (see 
page 353)
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= the percent of 
the overshoot of the selected 
waveform in NR3 format
:MEASure:PERiod 
[<source>] (see 
page 355)
:MEASure:PERiod? 
[<source>] (see 
page 355)
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>} for 
DSO models
<source> ::= {CHANnel<n> | 
DIGital<d> | FUNCtion | MATH | 
WMEMory<r>} for MSO models
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<return_value> ::= waveform 
period in seconds in NR3 format
:MEASure:PHASe 
[<source1>] 
[,<source2>] (see 
page 356)
:MEASure:PHASe? 
[<source1>] 
[,<source2>] (see 
page 356)
<source1,2> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= the phase 
angle value in degrees in NR3 
format
Table 70 :MEASure Commands Summary (continued)
Command Query Options and Query Returns