Technical data
338 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
n/a :MEASure:TVALue? 
<value>, 
[<slope>]<occurrence> 
[,<source>] (see 
page 365)
<value> ::= voltage level that 
the waveform must cross.
<slope> ::= direction of the 
waveform when <value> is crossed.
<occurrence> ::= transitions 
reported.
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>} for 
DSO models
<source> ::= {CHANnel<n> | 
DIGital<d> | FUNCtion | MATH | 
WMEMory<r>} for MSO models
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels 
- 1) in NR1 format
<return_value> ::= time in 
seconds of specified voltage 
crossing in NR3 format
:MEASure:VAMPlitude 
[<source>] (see 
page 367)
:MEASure:VAMPlitude? 
[<source>] (see 
page 367)
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= the amplitude 
of the selected waveform in volts 
in NR3 format
:MEASure:VAVerage 
[<interval>][,][<sour
ce>] (see page 368)
:MEASure:VAVerage? 
[<interval>][,][<sour
ce>] (see page 368)
<interval> ::= {CYCLe | DISPlay}
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= calculated 
average voltage in NR3 format
:MEASure:VBASe 
[<source>] (see 
page 369)
:MEASure:VBASe? 
[<source>] (see 
page 369)
<source> ::= {CHANnel<n> | 
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) 
in NR1 format
<r> ::= 1-2 in NR1 format
<base_voltage> ::= voltage at the 
base of the selected waveform in 
NR3 format
Table 70 :MEASure Commands Summary (continued)
Command Query Options and Query Returns