:MEASure:DELay
 (see page 790)
Command Syntax
:MEASure:DELay [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:DELay command places the instrument in the continuous 
measurement mode and starts a delay measurement.
The measurement is taken as:
delay = t(<edge spec 2>) -  t(<edge spec 1>)
where the <edge spec> definitions are set by the :MEASure:DEFine 
command
Query Syntax
:MEASure:DELay? [<source1>][,<source2>]
The :MEASure:DELay? query measures and returns the delay between 
source1 and source2. The delay measurement is made from the 
user- defined slope and edge count of the signal connected to source1, to 
the defined slope and edge count of the signal connected to source2. Delay 
measurement slope and edge parameters are selected using the 
:MEASure:DEFine command.
Also in the :MEASure:DEFine command, you can set upper, middle, and 
lower threshold values. It is the middle threshold value that is used when 
performing the delay query. The standard upper, middle, and lower 
measurement thresholds are 90%, 50%, and 10% values between Vbase and 
NOTE
The :MEASure:DELay command and the front-panel delay measurement differ from the 
:MEASure:DELay? query.
The delay command or front-panel measurement run the delay measurement in auto-edge 
select mode. In this mode, you can select the edge polarity, but the instrument will select 
the edges that will make the best possible delay measurement. The source1 edge chosen 
will be the edge that meets the polarity specified and is closest to the trigger reference 
point. The source2 edge selected will be that edge of the specified polarity that gives the 
first of the following criteria:
•
The smallest positive delay value that is less than source1 period.
• The smallest negative delay that is less than source1 period.
• The smallest absolute value of delay.
The :MEASure:DELay? query will make the measurement using the edges specified by the 