Technical data
:MEASure Commands 20
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 371
:MEASure:VMIN
 (see page 790)
Command Syntax
:MEASure:VMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VMIN command installs a screen measurement and starts a 
minimum vertical value measurement. If the optional source parameter is 
specified, the current source is modified.
Query Syntax
:MEASure:VMIN? [<source>]
The :MEASure:VMIN? query measures and outputs the minimum vertical 
value present on the selected waveform.
Return Format
<value><NL>
<value> ::= minimum vertical value of the selected waveform in
NR3 format
See Also • "Introduction to :MEASure Commands" on page 340
• ":MEASure:SOURce"  on page 361
• ":MEASure:VBASe"  on page 369
• ":MEASure:VMAX"  on page 370
• ":MEASure:VPP"  on page 372