Technical data
:MTESt Commands 21
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 383
:MTESt:AMASk:CREate
 (see page 790)
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask 
around the current selected channel, using the tolerance parameters 
defined by the :MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and 
:MTESt:AMASk:UNITs commands. The mask only encompasses the portion 
of the waveform visible on the display, so you must ensure that the 
waveform is acquired and displayed consistently to obtain repeatable 
results.
The :MTESt:SOURce command selects the channel and should be set before 
using this command.
See Also • "Introduction to :MTESt Commands" on page 379
• ":MTESt:AMASk:XDELta"  on page 386
• ":MTESt:AMASk:YDELta"  on page 387
• ":MTESt:AMASk:UNITs"  on page 385
• ":MTESt:AMASk:SOURce"  on page 384
• ":MTESt:SOURce"  on page 409
Example Code • "Example Code" on page 379