Technical data
:SAVE Commands 24
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 443
:SAVE:WAVeform:SEGMented
 (see page 790)
Command Syntax
:SAVE:WAVeform:SEGMented <option>
<option> ::= {ALL | CURRent}
When segmented memory is used for acquisitions, the 
:SAVE:WAVeform:SEGMented command specifies which segments are 
included when the waveform is saved:
• ALL — all acquired segments are saved.
• CURRent — only the currently selected segment is saved.
Query Syntax
:SAVE:WAVeform:SEGMented?
The :SAVE:WAVeform:SEGMented? query returns the current segmented 
waveform save option setting.
Return Format
<option><NL>
<option> ::= {ALL | CURR}
See Also • "Introduction to :SAVE Commands" on page 426
• ":SAVE:WAVeform[:STARt]"  on page 439
• ":SAVE:WAVeform:FORMat"  on page 440
• ":SAVE:WAVeform:LENGth"  on page 441