Technical data
:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 477
:SBUS<n>:IIC:TRIGger[:TYPE]
 (see page 790)
Command Syntax
:SBUS<n>:IIC:TRIGger[:TYPE] <value>
<value> ::= {STARt | STOP | READ7 | READEprom | WRITe7 | WRITe10
| NACKnowledge | ANACk | R7Data2 | W7Data2 | RESTart}
The :SBUS<n>:IIC:TRIGger[:TYPE] command sets the IIC trigger type:
• STARt — Start condition.
• STOP — Stop condition.
• READ7 — 7- bit address frame containing 
(Start:Address7:Read:Ack:Data). The value READ is also accepted for 
READ7.
• R7Data2 — 7- bit address frame containing 
(Start:Address7:Read:Ack:Data:Ack:Data2).
• READEprom — EEPROM data read.
• WRITe7 — 7- bit address frame containing 
(Start:Address7:Write:Ack:Data). The value WRITe is also accepted for 
WRITe7.
• W7Data2 — 7- bit address frame containing 
(Start:Address7:Write:Ack:Data:Ack:Data2).
• WRITe10 — 10- bit address frame containing (Start:Address 
byte1:Write:Ack:Address byte 2:Data).
• NACKnowledge — Missing acknowledge.
• ANACk — Address with no acknowledge.
• RESTart — Another start condition occurs before a stop condition.
Query Syntax
:SBUS<n>:IIC:TRIGger[:TYPE]?
The :SBUS<n>:IIC:TRIGger[:TYPE]? query returns the current IIC trigger 
type value.
Return Format
<value><NL>
<value> ::= {STAR | STOP | READ7 | READE | WRIT7 | WRIT10 | NACK | ANAC
| R7D2 | W7D2 | REST}
See Also • "Introduction to :TRIGger Commands" on page 589
• ":TRIGger:MODE"  on page 598
NOTE
The short form of READ7 (READ7), READEprom (READE), WRITe7 (WRIT7), and WRITe10 
(WRIT10) do not follow the defined Long Form to Short Form Truncation Rules (see 
page 792).