Technical data
486 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:LIN:SYNCbreak
 (see page 790)
Command Syntax
:SBUS<n>:LIN:SYNCbreak <value>
<value> ::= integer = {11 | 12 | 13}
The :SBUS<n>:LIN:SYNCbreak command sets the length of the LIN sync 
break to be greater than or equal to 11, 12, or 13 clock lengths. The sync 
break is the idle period in the bus activity at the beginning of each packet 
that distinguishes one information packet from the previous one.
Query Syntax
:SBUS<n>:LIN:SYNCbreak?
The :SBUS<n>:LIN:SYNCbreak? query returns the current LIN sync break 
setting.
Return Format
<value><NL>
<value> ::= {11 | 12 | 13}
See Also • "Introduction to :TRIGger Commands" on page 589
• ":TRIGger:MODE"  on page 598
• ":SBUS<n>:LIN:SOURce"  on page 484