Technical data
:SYSTem Commands 27
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 567
:SYSTem:ERRor
 (see page 790)
Query Syntax
:SYSTem:ERRor?
The :SYSTem:ERRor? query outputs the next error number and text from 
the error queue. The instrument has an error queue that is 30 errors deep 
and operates on a first- in, first- out basis. Repeatedly sending the 
:SYSTem:ERRor? query returns the errors in the order that they occurred 
until the queue is empty. Any further queries then return zero until 
another error occurs.
Return Format
<error number>,<error string><NL>
<error number> ::= an integer error code in NR1 format
<error string> ::= quoted ASCII string containing the error message
Error messages are listed in Chapter 34, “Error Messages,” starting on 
page 751.
See Also • "Introduction to :SYSTem Commands" on page 564
• "*ESR (Standard Event Status Register)" on page 128
• "*CLS (Clear Status)" on page 125