Technical data
s1
577
Agilent InfiniiVision 2000 X-Series Oscilloscopes 
Programmer's Guide
28
:TIMebase Commands
Control all horizontal sweep functions. See "Introduction to :TIMebase 
Commands"  on page 578.
Table 88 :TIMebase Commands Summary
Command Query Options and Query Returns
:TIMebase:MODE 
<value> (see 
page 579)
:TIMebase:MODE? (see 
page 579)
<value> ::= {MAIN | WINDow | XY | 
ROLL}
:TIMebase:POSition 
<pos> (see page 580)
:TIMebase:POSition? 
(see page 580)
<pos> ::= time from the trigger 
event to the display reference 
point in NR3 format
:TIMebase:RANGe 
<range_value> (see 
page 581)
:TIMebase:RANGe? (see 
page 581)
<range_value> ::= time for 10 div 
in seconds in NR3 format
:TIMebase:REFerence 
{LEFT | CENTer | 
RIGHt} (see page 582)
:TIMebase:REFerence? 
(see page 582)
<return_value> ::= {LEFT | CENTer 
| RIGHt}
:TIMebase:SCALe 
<scale_value> (see 
page 583)
:TIMebase:SCALe? (see 
page 583)
<scale_value> ::= time/div in 
seconds in NR3 format
:TIMebase:VERNier {{0 
| OFF} | {1 | ON}} 
(see page 584)
:TIMebase:VERNier? 
(see page 584)
{0 | 1}
:TIMebase:WINDow:POSi
tion <pos> (see 
page 585)
:TIMebase:WINDow:POSi
tion? (see page 585)
<pos> ::= time from the trigger 
event to the zoomed view 
reference point in NR3 format
:TIMebase:WINDow:RANG
e <range_value> (see 
page 586)
:TIMebase:WINDow:RANG
e? (see page 586)
<range value> ::= range value in 
seconds in NR3 format for the 
zoomed window
:TIMebase:WINDow:SCAL
e <scale_value> (see 
page 587)
:TIMebase:WINDow:SCAL
e? (see page 587)
<scale_value> ::= scale value in 
seconds in NR3 format for the 
zoomed window