Technical data
:TIMebase Commands 28
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 587
:TIMebase:WINDow:SCALe
 (see page 790)
Command Syntax
:TIMebase:WINDow:SCALe <scale_value>
<scale_value> ::= scale value in seconds in NR3 format
The :TIMebase:WINDow:SCALe command sets the zoomed (delayed) 
window horizontal scale (seconds/division). The main sweep scale 
determines the range for this command. The maximum value is one half of 
the :TIMebase:SCALe value.
Query Syntax
:TIMebase:WINDow:SCALe?
The :TIMebase:WINDow:SCALe? query returns the current zoomed window 
scale setting.
Return Format
<scale_value><NL>
<scale_value> ::= current seconds per division for the zoomed window
See Also • "Introduction to :TIMebase Commands" on page 578
• ":TIMebase:RANGe"  on page 581
• ":TIMebase:POSition"  on page 580
• ":TIMebase:SCALe"  on page 583
• ":TIMebase:WINDow:RANGe"  on page 586