:TRIGger:HFReject
 (see page 790)
Command Syntax
:TRIGger:HFReject <value>
<value> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:HFReject command turns the high frequency reject filter off 
and on. The high frequency reject filter adds a 50 kHz low- pass filter in 
the trigger path to remove high frequency components from the trigger 
waveform. Use this filter to remove high- frequency noise, such as AM or 
FM broadcast stations, from the trigger path.
Query Syntax
:TRIGger:HFReject?
The :TRIGger:HFReject? query returns the current high frequency reject 
filter mode.
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also • "Introduction to :TRIGger Commands" on page 589
• ":TRIGger[:EDGE]:REJect"  on page 604