Technical data
Getting Started 3
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 61
Other Ways of Sending Commands
Standard Commands for Programmable Instrumentation (SCPI) can also be 
sent via a Telnet socket or through the Browser Web Control:
• "Telnet Sockets" on page 61
• "Sending SCPI Commands Using Browser Web Control" on page 61
Telnet Sockets
The following information is provided for programmers who wish to 
control the oscilloscope with SCPI commands in a Telnet session.
To connect to the oscilloscope via a telnet socket, issue the following 
command:
telnet <hostname> 5024
where <hostname> is the hostname of the oscilloscope. This will give you a 
command line with prompt.
For a command line without a prompt, use port 5025. For example:
telnet <hostname> 5025
Sending SCPI Commands Using Browser Web Control
To send SCPI commands using the Browser Web Control feature, establish 
a connection to the oscilloscope via LAN as described in the InfiniiVision 
2000 X- Series Oscilloscopes User's Guide. When you make the connection 
to the oscilloscope via LAN and the instrument's welcome page is 
displayed, select the Browser Web Control tab, then select the Remote 
Programming link.