Technical data
:WAVeform Commands 30
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 639
:WAVeform:FORMat
 (see page 790)
Command Syntax
:WAVeform:FORMat <value>
<value> ::= {WORD | BYTE | ASCii}
The :WAVeform:FORMat command sets the data transmission mode for 
waveform data points. This command controls how the data is formatted 
when sent from the oscilloscope.
• ASCii formatted data converts the internal integer data values to real 
Y- axis values. Values are transferred as ASCii digits in floating point 
notation, separated by commas.
ASCII formatted data is transferred ASCii text.
• WORD formatted data transfers 16-bit data as two bytes. The 
:WAVeform:BYTeorder command can be used to specify whether the 
upper or lower byte is transmitted first. The default (no command sent) 
is that the upper byte transmitted first.
• BYTE formatted data is transferred as 8- bit bytes.
When the :WAVeform:SOURce is the serial decode bus SBUS1, ASCii is the 
only waveform format allowed.
When the :WAVeform:SOURce is one of the digital channel buses (BUS1 or 
BUS2), ASCii and WORD are the only waveform formats allowed.
Query Syntax
:WAVeform:FORMat?
The :WAVeform:FORMat query returns the current output format for the 
transfer of waveform data.
Return Format
<value><NL>
<value> ::= {WORD | BYTE | ASC}
See Also • "Introduction to :WAVeform Commands" on page 629
• ":WAVeform:BYTeorder"  on page 635
• ":WAVeform:SOURce"  on page 649
• ":WAVeform:DATA"  on page 637
• ":WAVeform:PREamble"  on page 644
Example Code • "Example Code" on page 650