Technical data
:WAVeform Commands 30
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 647
:WAVeform:SEGMented:COUNt
 (see page 790)
Query Syntax
:WAVeform:SEGMented:COUNt?
The :WAVeform:SEGMented:COUNt query returns the number of memory 
segments in the acquired data. You can use the 
:WAVeform:SEGMented:COUNt? query while segments are being acquired 
(although :DIGitize blocks subsequent queries until the full segmented 
acquisition is complete).
The segmented memory acquisition mode is enabled with the 
:ACQuire:MODE command. The number of segments to acquire is set using 
the :ACQuire:SEGMented:COUNt command, and data is acquired using the 
:DIGitize, :SINGle, or :RUN commands.
Return Format
<count> ::= an integer from 2 to 1000 in NR1 format (count set by
:ACQuire:SEGMented:COUNt).
See Also • ":ACQuire:MODE"  on page 187
• ":ACQuire:SEGMented:COUNt"  on page 190
• ":DIGitize"  on page 159
• ":SINGle"  on page 178
• ":RUN"  on page 176
• "Introduction to :WAVeform Commands" on page 629
Example Code • "Example Code" on page 191
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.