Technical data
672 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
31 :WGEN Commands
:WGEN:MODulation:AM:DEPTh
 (see page 790)
Command Syntax
:WGEN:MODulation:AM:DEPTh <percent>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN:MODulation:AM:DEPTh command specifies the amount of 
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used 
by the modulation. For example, a depth setting of 80% causes the output 
amplitude to vary from 10% to 90% (90% – 10% = 80%) of the original 
amplitude as the modulating signal goes from its minimum to maximum 
amplitude.
Query Syntax
:WGEN:MODulation:AM:DEPTh?
The :WGEN:MODulation:AM:DEPTh? query returns the AM depth 
percentage setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also • ":WGEN:MODulation:AM:FREQuency"  on page 673
• ":WGEN:MODulation:FM:DEViation"  on page 674
• ":WGEN:MODulation:FM:FREQuency"  on page 675
• ":WGEN:MODulation:FSKey:FREQuency"  on page 676
• ":WGEN:MODulation:FSKey:RATE"  on page 677
• ":WGEN:MODulation:FUNCtion"  on page 678
• ":WGEN:MODulation:FUNCtion:RAMP:SYMMetry"  on page 679
• ":WGEN:MODulation:STATe"  on page 681
• ":WGEN:MODulation:TYPE"  on page 682