Technical data
680 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
31 :WGEN Commands
:WGEN:MODulation:NOISe
 (see page 790)
Command Syntax
:WGEN:MODulation:NOISe <percent>
<percent> ::= 0 to 100
The :WGEN:MODulation:NOISe command adds noise to the currently 
selected signal. The sum of the amplitude between the original signal and 
injected noise is limited to the regular amplitude limit (for example, 5 Vpp 
in 1 MOhm), so the range for <percent> varies according to current 
amplitude.
Note that adding noise affects edge triggering on the waveform generator 
source as well as the waveform generator sync pulse output signal (which 
can be sent to TRIG OUT). This is because the trigger comparator is 
located after the noise source.
Query Syntax
:WGEN:MODulation:NOISe?
The :WGEN:MODulation:NOISe query returns the percent of added noise.
Return Format
<percent><NL>
<percent> ::= 0 to 100
See Also • ":WGEN:FUNCtion"  on page 667