:WGEN:VOLTage:OFFSet
 (see page 790)
Command Syntax
:WGEN:VOLTage:OFFSet <offset>
<offset> ::= offset in volts in NR3 format
The :WGEN:VOLTage:OFFSet command specifies the waveform's offset 
voltage or the DC level. Use the :WGEN:VOLTage command to specify the 
amplitude.
You can also specify the amplitude and offset indirectly using the 
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example, 
an amplitude of 5 V and an offset of 1 V is the same as a high- level 
voltage of 4 V and a low-level voltage of - 1 V.
Query Syntax
:WGEN:VOLTage:OFFSet?
The :WGEN:VOLTage:OFFSet? query returns the currently specified 
waveform offset voltage.
Return Format
<offset><NL>
<offset> ::= offset in volts in NR3 format
See Also • "Introduction to :WGEN Commands" on page 665
• ":WGEN:FUNCtion"  on page 667
• ":WGEN:VOLTage"  on page 688
• ":WGEN:VOLTage:HIGH"  on page 689
• ":WGEN:VOLTage:LOW"  on page 690